|
Home
» Temperature rise |
Temperature Rise |
«
Back |
|
|
|
-
Requirements relating to heating are
based on an ambient air temperature of
25.C [17.F). A temperature test may be
conducted at any ambient air temperature
and the variation from 25.C is added to
or subtracted from the observed
temperature reading.
-
Other than in those cases where it is
specifically stated that temperature
determinations are to be made
by the change-of-resistance method,
temperatures are to be measured by means
of thermocouples. A
thermocouple-measured temperature is to
be considered constant if three
successive readings, taken at intervals
of 10 percent of the previously elapsed
duration of the test (but at not less
than 5-minute intervals),
indicate no change. The junction of the
thermocouple is to be secured in contact
with the point on the surface at which
the temperature is to be measured. The
thermocouple is to consist of wires not
larger than No. 24 AWG (0.21 mm2
).
Exception No.1:
In here the thermocouple is used to
measure temperatures of electrically
live points, electrical insulation having
a maximum thickness of O.028 inch (0.71
mm) may be located between the
thermocouple and the live points.
|
|
|
|
Exception
No.2:
The coil temperature is to be determined by
the change-of-resistance method.
Exception No.3:
The ambient temperature may be determined by
a thermometer.
-
When
thermocouples are used in the
determination of temperatures in
connection with the heating of
electrical devices, it is standard practice
to use thermocouples consisting of No.
30AWG (0.05 mm2) iron and
constantan wires and a potentiometer-type
indicating instrument. Such equipment is
to be used whenever
referee temperature measurements are
necessary. The thermocouple wire is to
comply with the requirements for special
thermocouples as specified in the table
of limits of error of thermocouples in
Temperature Measurement Thermocouples,
ANSI MC96.1-1982.
-
The
temperature rise of a copper or aluminum
winding is to be determined by the
change-of-resistance
method using the following formula (windings
are to be at room temperature at the
start of the test):

in which:
t
is the temperature rise,
R
is the resistance of the coil at the end
of the test,
r
is the resistance of the coil at the
beginning of the test,
k
is 234.5 for copper and 225.0 for
aluminum,
t1
is the room temperature in degrees C at
the beginning of the test, and
t2
is the room temperature in degrees C at
the end of the test.
|
|